ASTM F1829 - 17 - Standard Test Method for Static Evaluation of Anatomic Glenoid Locking Mechanism in Shear

ASTM F1829 - 17

Standard Test Method for Static Evaluation of Anatomic Glenoid Locking Mechanism in Shear

Status : Current   Published : December 2017

Format
PDF

Format
HARDCOPY



1.1 This test method covers a method for determining the static shear disassembly force of modular anatomic glenoid components used in anatomic total shoulder arthroplasty prostheses.

1.2 Although the methodology described does not replicate all physiological force conditions, it is a means of in vitro comparison of modular anatomic glenoid component designs and the strength of the retention mechanism between the articular insert and glenoid backing under the stated test conditions.

1.3 This test method covers modular glenoid components comprised of a separate articular insert and backing. The insert and backing can be fabricated from any combination of the following materials: metal alloys, polymeric materials, composite materials.

1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5  This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

1.6  This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.




Standard NumberASTM F1829 - 17
TitleStandard Test Method for Static Evaluation of Anatomic Glenoid Locking Mechanism in Shear
StatusCurrent
Publication Date01 December 2017
Normative References(Required to achieve compliance to this standard)No other standards are normatively referenced
Informative References(Provided for Information)No other standards are informatively referenced
Descriptors Arthroplasty, Disassembly, Glenoid, Modular, Orthopaedic medical devices, Shoulder arthroplasty
ICS11.040.40
PublisherASTM
FormatA4
DeliveryYes
Pages4
File Size115 KB
Price£37.00


 Your basket
Your basket is empty

Multi-user access to over 3,500 medical device standards, regulations, expert commentaries and other documents


Tracked Changes

Understand the changes made to a standard with our new Tracked Changes version


Develop a PAS

Develop a fast-track standardization document in 9-12 months


Worldwide Standards
We can source any standard from anywhere in the world