ASTM F2777 - 16 - Standard Test Method for Evaluating Knee Bearing (Tibial Insert) Endurance and Deformation Under High Flexion

ASTM F2777 - 16

Standard Test Method for Evaluating Knee Bearing (Tibial Insert) Endurance and Deformation Under High Flexion

Status : Current   Published : July 2016

Format
PDF

Format
HARDCOPY



1.1 This standard specifies a test method for determining the endurance properties and deformation, under specified laboratory conditions, of ultra high molecular weight polyethylene (UHMWPE) tibial bearing components used in bicompartmental or tricompartmental knee prosthesis designs.

1.2 This test method is intended to simulate near posterior edge loading similar to the type of loading that would occur during high flexion motions such as squatting or kneeling.

1.3 Although the methodology described attempts to identify physiological orientations and loading conditions, the interpretation of results is limited to an in vitro comparison between knee prosthesis designs and their ability to resist deformation and fracture under stated test conditions.

1.4 This test method applies to bearing components manufactured from UHMWPE.

1.5 This test method could be adapted to address unicompartmental total knee replacement (TKR) systems, provided that the designs of the unicompartmental systems have sufficient constraint to allow use of this test method. This test method does not include instructions for testing two unicompartmental knees as a bicompartmental system.

1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.7  This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.




Standard NumberASTM F2777 - 16
TitleStandard Test Method for Evaluating Knee Bearing (Tibial Insert) Endurance and Deformation Under High Flexion
StatusCurrent
Publication Date01 July 2016
Normative References(Required to achieve compliance to this standard)No other standards are normatively referenced
Informative References(Provided for Information)No other standards are informatively referenced
Descriptors Deformation, Fatigue, Fracture, High flexion, Knee, Mobile bearing, Rotation, TKR (total knee replacement), UHMWPE (ultra high molecular weight polyethylene)
ICS11.040.40
PublisherASTM
FormatA4
DeliveryYes
Pages8
File Size290 KB
Price£40.00


 Your basket
Your basket is empty

Multi-user access to over 3,500 medical device standards, regulations, expert commentaries and other documents


Develop a PAS

Develop a fast-track standardization document in 9-12 months


Tracked Changes

Understand the changes made to a standard with our new Tracked Changes version


BSOL

The faster, easier way to work with standards