1.1 This standard will measure the effects on the accuracy of computer assisted surgery (CAS) systems of the environmental influences caused by equipment utilized for bone preparation during the intended clinical application for the system. The environmental vibration effect covered in this standard will include mechanical vibration from: Cutting saw (sagittal or reciprocating), Burrs, drills and impact loading. The change in accuracy from detaching and re-attaching, or disturbing a restrained connection that does not by design require repeating the registration process of a reference base will also be measured.

1.2 It should be noted that one system may need to undergo multiple iterations (one for each clinical application) of this standard to document its accuracy during different clinical applications since each procedure may have different exposure to outside forces given the surgical procedure variability from one procedure to the next.

1.3 All units of measure will be reported as millimeters for this standard.

1.4  This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.




Standard NumberASTM F3107 - 14
TitleStandard Test Method for Measuring Accuracy after Mechanical Disturbances on Reference Frames of Computer Assisted Surgery Systems
StatusCurrent
Publication Date12 January 2014
Normative References(Required to achieve compliance to this standard)No other standards are normatively referenced
Informative References(Provided for Information)No other standards are informatively referenced
Descriptors Computer assisted surgery, Computer navigation, Imageless guided surgery, Infrared tracking system, Optical tracking
ICS11.040.01
PublisherASTM
FormatA4
DeliveryYes
Pages5
File Size96 KB
Price£40.00


 Your basket
Your basket is empty

Multi-user access to over 3,500 medical device standards, regulations, expert commentaries and other documents


Tracked Changes

Understand the changes made to a standard with our new Tracked Changes version


BSOL

The faster, easier way to work with standards


Worldwide Standards
We can source any standard from anywhere in the world